MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER TIEMANN MODEL COUDE; RED RUBBER CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER TIEMANN MODEL COUDE; RED RUBBER CATHETER Back to Search Results
Model Number 0102L12
Medical Device Problem Code Failure to Infuse (2340)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 05/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the catheter allegedly inflated in the middle of the shaft, as opposed to the tip.The catheter fell out of the patient after 5 minutes.A second unit was used for the patient.
 
Event or Problem Description
It was reported that the catheter allegedly inflated in the middle of the shaft, as opposed to the tip.The catheter fell out of the patient after 5 minutes.A second unit was used for the patient.
 
Additional Manufacturer Narrative
The reported event was confirmed as manufacturing related.A blister was observed on the irrigation lumen at the midpoint on the shaft.The device was inflated with 10ml of water.The force of the inflation liquid injected tore through thin lumen walls and burst out of the blister.Per specification, blisters are considered manufacturing defects.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.' correction: mfr site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER TIEMANN MODEL COUDE
Common Device Name
RED RUBBER CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7550734
Report Number1018233-2018-01934
Device Sequence Number16187456
Product Code EZC
UDI-Device Identifier00801741018107
UDI-Public(01)00801741018107
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2021
Device Model Number0102L12
Device Catalogue Number0102L12
Device Lot NumberNGAZ2591
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2018
Initial Date Received by Manufacturer 05/07/2018
Supplement Date Received by Manufacturer08/23/2018
Initial Report FDA Received Date05/29/2018
Supplement Report FDA Received Date09/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No