MEDTRONIC SINGAPORE OPERATIONS CAPSURE SENSE; ELECTRODE, PACEMAKER, PERMANENT Back to Search Results
Model Number 407452
Medical Device Problem Codes High Capture Threshold (3266); Unstable Capture Threshold (3269)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/03/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the right ventricular (rv) lead had high and variable thresholds.When the lead was programmed unipolar, the patient could feel the pacing.The lead remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was further reported that programming changes were made to adjust the output and turn off adaptive capture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE SENSE
Common Device Name
ELECTRODE, PACEMAKER, PERMANENT
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SG  486056
MDR Report Key7551642
Report Number3008973940-2018-01231
Device Sequence Number274845
Product Code DTB
UDI-Device Identifier00643169448940
UDI-Public00643169448940
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P830061
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/07/2019
Device Model Number407452
Device Catalogue Number407452
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/03/2018
Supplement Date Received by Manufacturer06/05/2018
Initial Report FDA Received Date05/29/2018
Supplement Report FDA Received Date06/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ADDR01 IPG, 457445 LEAD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
-
-