| Model Number |
407452 |
| Medical Device Problem Codes |
High Capture Threshold (3266); Unstable Capture Threshold (3269)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
05/03/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead had high and variable thresholds.When the lead was programmed unipolar, the patient could feel the pacing.The lead remains in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was further reported that programming changes were made to adjust the output and turn off adaptive capture.
|
| |
|