• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. PROXIMATE*RL TITAN HEAVY WIRE; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. PROXIMATE*RL TITAN HEAVY WIRE; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number TLH60
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that during the colectomy, after fired, found the staples malformed when severing the colon.Suture was used to sew the wound.There were no adverse consequences to the patient.
 
Additional Manufacturer Narrative
(b)(4).Batch # p56p0x.Device evaluation: the analysis results showed that the tlh60 device arrived with no apparent damage with a reload loaded on the device.The reload was returned void of staples.The device was tested for functionality with a test reload and it cycled, fired, and all the staples formed as intended.The device fired without any difficulties and the staples were noted to have the proper b-formed shape.The event described could not be confirmed as the device performed without any difficulties noted.This report is not intended to deny that you experienced a problem with the device.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7552281
Report Number3005075853-2018-10233
Device Sequence Number11215866
Product Code GDW
UDI-Device Identifier20705036013232
UDI-Public20705036013232
Combination Product (Y/N)N
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/23/2022
Device Catalogue NumberTLH60
Device Lot NumberP4RW2L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2018
Initial Date Received by Manufacturer 05/04/2018
Supplement Date Received by Manufacturer06/07/2018
Initial Report FDA Received Date05/30/2018
Supplement Report FDA Received Date06/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-