• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-522NAP
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that they received compromised force sensor system alarm as well as prime rewind.The customer¿s blood glucose level was 187 mg/dl.Troubleshooting was performed for prime rewind.The customer also stated that the bottom of the pump was coming loose.The product will be returned for analysis.
 
Additional Manufacturer Narrative
Device received with broken off end cap.Unable to perform rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test or verify compromised force sensor system alarm due to broken off end cap.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAP
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7552455
Report Number3004209178-2018-75253
Device Sequence Number1818203
Product Code OYC
UDI-Device Identifier00613994402103
UDI-Public(01)00613994402103
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-522NAP
Device Catalogue NumberMMT-522NAP
Device Lot NumberA5522NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2018
Initial Date Received by Manufacturer 04/30/2018
Supplement Date Received by Manufacturer08/18/2018
Initial Report FDA Received Date05/30/2018
Supplement Report FDA Received Date09/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age49 YR
Patient Weight154
-
-