MPRI SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 977D260
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/07/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis of the epidural needle found the spoon of the needle had a sharp edge on the undercut side.Product id: neu_epiduralneedle, lot# serial# unknown, product type: accessory.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider (hcp) via a manufacturer representative.It was reported that the surgeon passed the lead through the end of the included curved tip needle.In the course of placing the lead, the surgeon attempted to slightly withdraw the lead and it got stuck in the tip of the needle.The surgeon had to remove the needle and forcefully free the lead from needle.Additional information was received from a manufacture representative (rep) on (b)(6) 2017.The rep stated that the cause remains unknown, so the item would be returned on 2017-nov-20.The lead, stylet, and epidural needle were received for analysis on (b)(6) 2017.There were no patient symptoms reported and no complications reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7552735
Report Number2649622-2018-08968
Device Sequence Number144014
Product Code LGW
UDI-Device Identifier00643169118249
UDI-Public00643169118249
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P840001
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/04/2021
Device Model Number977D260
Device Catalogue Number977D260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/01/2018
Supplement Date Received by Manufacturer11/06/2018
Initial Report FDA Received Date05/30/2018
Supplement Report FDA Received Date11/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age57 YR
-
-