| Model Number |
MNC20008 |
| Medical Device Problem Codes |
Device Operational Issue (2914); Difficult to Advance (2920); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of device history record including (b)(4) material inspection, in process & finished product inspection does not reveal any discrepancy relevant to the batch under investigation which confirms that there is no indication of a product related quality deficiency associated to this batch.Additionally, a review of the complaint history identified no other incidents from this lot.
|
| |
|
Event or Problem Description
|
|
A 2.0 x 8 mozec nc was advanced to a lesion in the rv marginal branch but it could not cross the lesion.Another ptca catheter (bsc - 2.0 x 8 nc emerge) used and the procedure completed.No issue to the patient noticed.
|
| |
|