| Model Number |
41701 |
| Medical Device Problem Codes |
Device Operates Differently Than Expected (2913); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that after the stylet was removed from the catheter, it was unable to be reinserted.It was questioned if the catheter was blocked.The stylet that was packaged with the catheter was the same stylet being used.A different catheter was used to complete the procedure with no issues.
|
| |
|
Additional Manufacturer Narrative
|
|
Only the ventricular catheter was returned.The catheter was patent and met the requirements for leak testing.All catheters are 100% inspected at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|