Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Blood Loss (2597); No Known Impact Or Consequence To Patient (2692)
Date of Event 05/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.The actual device has been returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.For this reason, evaluation code has been referenced in the conclusions.A review of the device history record of the product code/lot# combination was conducted with no findings.This report is for the fourth device that was used on the same patient.See mdr's 3013394970-2018-00305, 3013394970-2018-00306, 3013394970-2018-00307 for the other three reported events.
 
Event or Problem Description
The user facility reported they used four different angio-seal sts 8fr devices.The customer had issues with the collagen/bypass tube staying in place.They did not attempt to place the first three devices in the patient.The customer said the device was coming apart near the collage/bypass tube.The customer used the fourth device on the patient.Upon deployment the customer noticed that the device was coming apart and did not have good hemostasis.The customer held pressure on the patient and that controlled the bleeding.Estimated blood loss was less than 250cc.The procedure outcome was successful.Manual compression was used to close the artery.
 
Additional Manufacturer Narrative
One 8fr angio-seal sts plus was returned for evaluation.One carrier tube assembly was returned.The bypass tube was not returned.The collagen within the carrier tube was slightly swollen.The anchor was exposed due to detachment of the bypass tube.There was no damage noted on the mating wings of the carrier tube hub.The aluminum pouch was not returned.The following measurements were carried out: carrier tube outer diameter = 0.1019" and tamper tube outer diameter = 0.0794".All measurements were within the specifications defined in the engineering drawings active at the time of manufacturing.There is no evidence that this event was related to a device defect or malfunction.Based on evaluation of the returned carrier tube assembly, no anomalies could be detected during microscopic and dimensional inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8F ANGIO-SEAL STS PLUS VASCULAR CLOSURE DEVICE, US
Common Device Name
DEVICE, HEMOSTASIS, VASCULAR
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
2101 cottontail ln.
somerset NJ 08873
MDR Report Key7553879
Report Number3013394970-2018-00308
Device Sequence Number1277772
Product Code MGB
UDI-Device Identifier00389701011851
UDI-Public00389701011851
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P930038
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2019
Device Model NumberN/A
Device Catalogue Number610121
Device Lot Number06085475
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/17/2018
Supplement Date Received by Manufacturer07/11/2018
Initial Report FDA Received Date05/30/2018
Supplement Report FDA Received Date07/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-