• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; COBRA GRASPER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; COBRA GRASPER Back to Search Results
Model Number 420190-08
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the customer reported failure.The instrument was found to have a broken pitch cable at the distal clevis hub.The broken cable segment that contains the crimp was still installed in the clevis.The clevis did not exhibit any damage or wear marks.Other cables at the wrist were not damaged.Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to this reported event.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis suggests that if the malfunction were to recur it could cause or contribute to an adverse event.
 
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, the customer noted that the cobra grasper instrument had a broken wire.There was no report of fragment(s) falling inside the patient, patient harm, adverse outcome or injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Common Device Name
COBRA GRASPER
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
950 kifer rd.
sunnyvale
MDR Report Key7553997
Report Number2955842-2018-10222
Device Sequence Number178912
Product Code NAY
UDI-Device Identifier00886874111598
UDI-Public(01)00886874111598(10)N10170919
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K050369
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420190-08
Device Lot NumberN10170919 772
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2018
Initial Report FDA Received Date05/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-