| Model Number |
977D260 |
| Medical Device Problem Codes |
Break (1069); Difficult to Remove (1528); Device Damaged Prior to Receipt by User (1570); Device Operates Differently Than Expected (2913); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis of the epidural needle found the spoon of the needle had a sharp edge on the undercut side.Concomitant medical products: product id: neu_epidural needle, lot# unknown, product type: accessory.Product id: neu_epidural needle, lot# unknown, product type: accessory.Information references the main component of the system.Other relevant device(s) are: product id: neu_epidural needle, serial/lot #: unknown, ubd: , udi#: ; product id: neu_epidural needle, serial/lot #: unknown, ubd: , udi#: this report is a correction.Regulatory report 3007566237-2018-00433 captured the initial report.No further regulatory reports will be submitted under report number 3007566237-2018-00433.If additional information is received it will be captured in this regulatory report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a manufacturer representative concerning a trial patient.An out of box failure was reported.It was reported that healthcare provider (hcp) complained the curved tip needles provided in leads kits were poor quality.When inserting the lead into the needle, it is not moving smoothly through the hub and once the electrodes exit the needles, it is extremely difficult to retract the lead back into the needle.There was some type of partial occlusion or burr or something like that.He also added that when an electrode is outside of the needle tip, it is difficult to retract and he has to redo the epidural.The hcp demonstrated it today for the representative and she observed it as well.The hcp commented "these needles are terrible." no patient symptoms were reported.No further complications were reported/anticipated.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to imdrf harmonization, some previously submitted device, method, result, and conclusion codes related to this event may have been updated.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|