• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ECLIPSE FILTER SYSTEM - FEMORAL; VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ECLIPSE FILTER SYSTEM - FEMORAL; VENA CAVA FILTER Back to Search Results
Catalog Number EC500J
Device Problems Migration or Expulsion of Device (1395); Malposition of Device (2616); Extrusion (2934); Patient Device Interaction Problem (4001)
Patient Problems Abdominal Pain (1685); No Consequences Or Impact To Patient (2199)
Event Date 04/20/2017
Event Type  Injury  
Manufacturer Narrative
Manufacturing review: as the lot number for the device was not provided, a manufacturing review could not be performed.Investigation summary: the device was not returned for evaluation.Images and medical records were not provided for review.Therefore, the investigation is inconclusive for filter perforation, tilt and migration post filter deployment as no objective evidence has been provided to confirm any alleged deficiency with the filter.Based upon the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported through the litigation process that some time post vena cava filter deployment the filter perforated, tilted, and migrated.There were no reported attempts made to retrieve the filter.The status of the patient is unknown.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was not returned for evaluation.Medical records were provided and reviewed.Approximately four years two months of post deployment, the patient presented with history of abdominal pain and a computerized tomography-abdomen without contrast was performed which showed that the filter was tilted to the right with its cone lying on the right lateral wall of the inferior vena cava possibly embedded within it.The legs of the filter had penetrated through the wall of the inferior vena cava into the pericaval/mesenteric fat.Therefore, the investigation is confirmed for filter tilt and perforation of the inferior vena cava (ivc).However, the investigation is inconclusive for filter migration.Based upon the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.H10: (expiration date: 10/2015).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed after being diagnosed with deep vein thrombosis/pulmonary embolism.At some time post filter deployment, it was alleged that the filter migrated, tilted, struts perforated and embedded in wall of the inferior vena cava.The device has not been removed and there were no reported attempts made to retrieve the filter.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECLIPSE FILTER SYSTEM - FEMORAL
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7555133
MDR Text Key109623600
Report Number2020394-2018-00755
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K093659
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup
Report Date 12/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/30/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEC500J
Device Lot NumberGFWI3176
Was Device Available for Evaluation? No
Date Manufacturer Received11/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
COUMADIN, NITROGLYCERIN, LOVENOX, CYMBALTA; LEVOTHYROXINE; LORTAB, NITROSTAT, VALIUM, AMITRIPTYLINE; METFORMIN, CARISOPRODOL, DIPYRIDAMOLE; SERTRALINE, CRESTOR, DUO NEB, METOPROLOL
Patient Outcome(s) Other;
Patient Age62 YR
Patient Weight102
-
-