| Model Number |
S9385108 |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of rebu1912 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported that a powerpicc was placed on (b)(6) 2018.It was stated that during insertion it was difficult to thread and the wire broke off.Post insertion x-ray showed a 3cm piece of wire in the pulmonary artery.At the time of report the wire was still in the patient.On (b)(6) 2018 - the facility reported that "the wire is in the pulmonary artery and is not causing harm at this time.It may interfere with care in the future though, i.E.No mris." the guidewire will not be removed as removal is too dangerous due to the location.It will stay in the patient indefinitely.The facility does not know if any trouble shooting measures were attempted or if there were issues with insertion.Patient has past history of cvc placements.
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a broken stylet was confirmed and the cause appeared to be use-related.The product returned for evaluation was one tls stylet.The sample was received inlaid through a t-lock assembly.Usage residues were observed on the sample.Several bends and kinks were observed along the length of the stylet.The distal tip was detached and was not returned for evaluation.Microscopic inspection of the distal end of the sample revealed kinking and plastic deformation at the break site.The break site exhibited an elliptical cross-section.The break occurred between two magnets.The deformation and kinks in the polyimide coating and the break characteristics were consistent with damage caused by kinking and pulling the stylet.The usage residues and event description indicated that kinking and pulling occurred during device use.A lot history review (lhr) of rebu1912 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported that a powerpicc was placed on (b)(6) 2018.It was stated that during insertion it was difficult to thread and the wire broke off.Post insertion x-ray showed a 3cm piece of wire in the pulmonary artery.At the time of report the wire was still in the patient.05/16/2018-the facility reported that "the wire is in the pulmonary artery and is not causing harm at this time.It may interfere with care in the future though, i.E.No mris." the guidewire will not be removed as removal is too dangerous due to the location.It will stay in the patient indefinitely.The facility does not know if any trouble shooting measures were attempted or if there were issues with insertion.Patient has past history of cvc placements.
|
| |
|