• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII OVAL RESURFACING PAT 32MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CMNTD, POLY/MTL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII OVAL RESURFACING PAT 32MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CMNTD, POLY/MTL/POLY Back to Search Results
Catalog Number 71421032
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Codes Pain (1994); Injury (2348)
Date of Event 06/21/2016
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent a revision surgery on the left knee due to pain and stiffness.Properative films revealed a possible loosening of the patellar implant and significant heterotopic ossification, both were confirmed during revision.Only the patellar component was revised.
 
Additional Manufacturer Narrative
The associated complaint device was not returned.A clinical evaluation was conducted and based on the limited information provided, the root cause of the patient's reported pain and mua cannot be concluded; however, the loose patella was likely a contributing factor for the pain as well as the left tka revision.The impact to the patient beyond the expected pain from the revision procedure itself cannot be concluded.No further medical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII OVAL RESURFACING PAT 32MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CMNTD, POLY/MTL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7555300
Report Number1020279-2018-01018
Device Sequence Number16473286
Product Code JWH
UDI-Device Identifier03596010485069
UDI-Public03596010485069
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Attorney
Type of Report Initial,Followup
Report Date (Section B) 08/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/16/2024
Device Catalogue Number71421032
Device Lot Number14FM13479
Initial Date Received by Manufacturer 05/29/2018
Supplement Date Received by Manufacturer05/29/2018
Initial Report FDA Received Date05/30/2018
Supplement Report FDA Received Date08/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age65 YR
-
-