Model Number DTBB1Q1
Medical Device Problem Code Defibrillation/Stimulation Problem (1573)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during the implant procedure, defibrillation threshold (dft) testing was performed with difficulty and required external rescue.After multiple attempts to convert ventricular fibrillation (vf) with 10j safety margin were not successful the cardiac resynchronization therapy defibrillator (crt-d) was removed and replaced with another device.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA QUAD S
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key7555346
Report Number3004209178-2018-12188
Device Sequence Number1277781
Product Code NIK
UDI-Device Identifier00643169530201
UDI-Public00643169530201
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/14/2018
Device Model NumberDTBB1Q1
Device Catalogue NumberDTBB1Q1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2018
Initial Date Received by Manufacturer 05/07/2018
Supplement Date Received by Manufacturer06/12/2018
Initial Report FDA Received Date05/30/2018
Supplement Report FDA Received Date06/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age79 YR
-
-