MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION BSM-1773A; BEDSIDE MONITOR, NK SPO2

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION BSM-1773A; BEDSIDE MONITOR, NK SPO2 Back to Search Results
Model Number BSM-1773A
Medical Device Problem Codes Thermal Decomposition of Device (1071); Device Operational Issue (2914); Compatibility Problem (2960)
Health Effect - Clinical Codes Abrasion (1689); Pressure Sore/Ulcer (2326); Injury (2348); Patient Problem/Medical Problem (2688)
Date of Event 04/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The nursing manager reported that an spo2 probe caused an open abrasion on the right foot of a neonate patient.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event or Problem Description
The nursing manager reported that an spo2 probe caused an open abrasion on the right foot of a neonate patient.
 
Additional Manufacturer Narrative
Manufacturer narrative: investigation result: this complaint is regarding spo2 probe causing injuries, not the malfunction of the monitoring device that the probe was connected to.The above incident was reported on (b)(6) 2018 during a go-live event.There are nine other incidents of injury from this go-live event.The issue is isolated to one customer: (b)(6).The following is the list of additional tickets related to this incident: (b)(4).In the nkc report (b)(4), it was reported that customer had placed the probe on the neonate patient for longer than the recommended maximum of 8 hours.The average attached time of the probe was 12 hours when the reported injury occurred.Customer had intended to keep the probe attached for 48 hours.Tl-052s is a disposable tape.The (b)(4) found that customer had reused the probe, which is not within the intended use of the probe.Operating manuals for g9 (csm-1000 series operator's manual, page 3-14) and bsm-1700 (operator's manual bsm-1700, page 123) posted warnings to change measurement sites for spo2 probes every 8 hours.Circulation condition should be checked by observing skin color at the measurement site and pulse waveforms.Extreme care should be taken when using probes on neonate or low birth weight infant with delicate skin.From an email by (b)(4), nk clinical applications specialist, on (b)(4) 2018, that during the go-live event, the customer stated the spo2 probes were lasting only three hours without them having to change the probe.This is alluding to the adhesive tape on the probe, although no clarification was provided.The situation became a huge expense to the unit and babies were having to stay longer based on desats that they felt may or may not be real due to the probe coming loose.Customer stated that the probe is stiff and hard which made to more difficult to get the sensors to touch the skin.The tape that came with these probes is very stretchy so they would put the probe on and if the tape was not tight enough the sensor would raise causing a false alarm.They would put the tape on tighter so that the sensor would not raise; this is when it would cause cuts to the babies' skin because the probe itself is hard.Based on the alleged spo2 inaccuracy and probe coming loose, the following queries were performed to see if these were recurring issues.A query for tickets related to "tl-052s" shows no other tickets than the ones listed above.A query for tickets related to "spo2 probe loose" shows no results.A query for tickets related to "spo2 accuracy" shows three incidents: ticket (b)(4) - not relevant.Customer reported artifacts on waveforms rather than an alleged spo2 inaccuracy.Ticket (b)(4) - not relevant.Customer asked clinical knowledge question rather than an alleged spo2 inaccuracy.O ticket 43761 - investigation is in progress.Limited information available.Root cause has not been determined.Query results show tl-052s spo2 probe incidents were isolated to (b)(6).There are no known spo2 probes coming loose.There is one other incident of alleged spo2 inaccuracy that is still under investigation.Nkc's investigation results did not find issues with the hardness of the sensor area and tape shape for tl-052s.Risk analysis was performed under (b)(4).It was judged that the risk evaluation was acceptable.From the above information, it is concluded that the root cause was due to customer not being familiar with nk spo2 technology and have misused the product.The risks associated was determined by nk clinical department: if the spo2 probe is causing pressure marks on the neonate this could a serious concern as the neonate is not able to report the pain from the injury.The reddened pressure mark could cause tissue damage as well as inaccurate spo2 readings.Inaccurate spo2 reading could lead to over and/or under treating neonates with oxygen as well as delay in treatment.As a correction, customer had elected to have devices with nk spo2 technology swapped out for nellcor technology.This issue does not affect any other customers.Update: on 4/12/2019, customer sent an excel sheet containing information regarding eleven (11) cases.Ten (10) of which were mentioned above.Nk tech support is in contact with customer.Pending more information from customer.
 
Additional Manufacturer Narrative
Additional narrative: it was reported that an infant has an open abrasion on the right foot.Hospital was using nk spo2 probes in this area and saw the cut once they removed the probe.Monitoring device: bsm-1773a sn not available.Service requested: troubleshooting.Service performed: troubleshooting.Investigation result: this is part of the collective incident reported from sparrow health system.(b)(4).Investigation from (b)(4): investigation result: this complaint is regarding spo2 probe causing injuries, not the malfunction of the monitoring device that the probe was connected to.The above incident was reported on 5/2/2018 during a go-live event.There are nine other incidents of injury from this go-live event.The issue is isolated to one customer: sparrow health system (account id (b)(4)).The following is the list of additional tickets related to this incident: (b)(4).In the nkc report (b)(4)_final_eg, it was reported that customer had placed the probe on the neonate patient for longer than the recommended maximum of 8 hours.The average attached time of the probe was 12 hours when the reported injury occurred.Customer had intended to keep the probe attached for 48 hours.Tl-052s is a disposable tape.The (b)(4)_final_eg found that customer had reused the probe, which is not within the intended use of the probe.Operating manuals for (csm-1000 series operator's manual, page 3-14) and (b)(4) (operator's manual bsm-1700, page 123) posted warnings to change measurement sites for spo2 probes every 8 hours.Circulation condition should be checked by observing skin color at the measurement site and pulse waveforms.Extreme care should be taken when using probes on neonate or low birth weight infant with delicate skin.From an email by (b)(6), nk clinical applications specialist, on (b)(6) 2018, that during the go-live event, the customer stated the spo2 probes were lasting only three hours without them having to change the probe.This is alluding to the adhesive tape on the probe, although no clarification was provided.The situation became a huge expense to the unit and babies were having to stay longer based on desats that they felt may or may not be real due to the probe coming loose.Customer stated that the probe is stiff and hard which made to more difficult to get the sensors to touch the skin.The tape that came with these probes is very stretchy so they would put the probe on and if the tape was not tight enough the sensor would raise causing a false alarm.They would put the tape on tighter so that the sensor would not raise; this is when it would cause cuts to the babies' skin because the probe itself is hard.Query results show tl-052s spo2 probe incidents were isolated to sparrow health system (account id (b)(4)).There are no known spo2 probes coming loose.There is one other incident of alleged spo2 inaccuracy that is still under investigation.Nkc's investigation results did not find issues with the hardness of the sensor area and tape shape for tl-052s.Risk analysis was performed under (b)(4).It was judged that the risk evaluation was acceptable.From the above information, it is concluded that the root cause was due to customer not being familiar with nk spo2 technology and have misused the product.The risks associated was determined by nk clinical department: if the spo2 probe is causing pressure marks on the neonate this could a serious concern as the neonate is not able to report the pain from the injury.The reddened pressure mark could cause tissue damage as well as inaccurate spo2 readings.Inaccurate spo2 reading could lead to over and/or under treating neonates with oxygen as well as delay in treatment.As a correction, customer had elected to have devices with nk spo2 technology swapped out for nellcor technology.This issue does not affect any other customers.Corrected information: date of event: date incorrectly notated.Should be (b)(6) 2018.Date uf/importer became aware of event: aware date on mdr initial should be 04/19/2018 not 05/03/2018.Approximate age of device: should be left blank, not 0 months as no serial # provided.Date received by manufacturer: should be 04/19/2018 not 05/30/2018.As listed on mdr initial report type of reportable event: serious injury should have been selected not malfunction on the mdr initial.Mdr follow up 002 has serious injury selected.Additional manufacturer narrative: the following fields are not applicable (n/a) to this report: lot # & expiration date (na for the bsm not for the spo2 probe) concomitant medical product.Spo2 probe information: model: tl-052s, lot: not provided.Expiration date: not provided.Unique identifier (udi) #: (b)(4).Is this a single-use device that was reprocessed and reused on a patient? no.Probes used in this incident were not returned the spo2 probe was used in conjunction with the bedside monitor.Approximate age of device for spo2 probe: probe lot #/expiration not provided.No age of probe available.Device manufacture date for spo2 probe: probe lot #/expiration not provided.No device manufacture date available.Labeled for single use? yes.Usage of device: reuse.The probe is labeled for single use but the customer re-used it anyway which resulted in a patient injury.
 
Event or Problem Description
The nursing manager reported that an spo2 probe caused an open abrasion on the right foot of a neonate patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BSM-1773A
Common Device Name
BEDSIDE MONITOR, NK SPO2
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, 161-8 560
JA  161-8560
MDR Report Key7555499
Report Number8030229-2018-00182
Device Sequence Number11269993
Product Code MHX
UDI-Device Identifier04931921111956
UDI-Public4931921111956
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K080342
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/06/2019
Report Date (Section F) 09/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBSM-1773A
Device Catalogue NumberBSM-1773A
Device Lot NumberNOT APPLICABLE
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date04/29/2019
Device Age0 MO
Event Location Hospital
Date Report to Manufacturer09/06/2019
Initial Date Received by Manufacturer 05/30/2018
Supplement Date Received by Manufacturer05/30/2018
04/29/2019
Initial Report FDA Received Date05/30/2018
Supplement Report FDA Received Date04/12/2019
09/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NOT APPLICABLE
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Date Report Sent to FDA09/06/2019
-
-