ANGIODYNAMICS ANGIODYNAMICS / VORTEX; PORT & CATHETER, IMPLANTED Back to Search Results
Catalog Number H787LVTX52130
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Pulmonary Embolism (1498)
Date of Event 02/03/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
In lieu of a reported lot number, a ship history report (shr) was generated for item number ((b)(4)) in order to determine the last three lots shipped to the reporting hospital.The shr indicated the reporting hospital only received 2 devices {one lot} of this item number since (b)(6) 2012; shipped to the hospital on (b)(6) 2015.This is inconsistent with information that there are 7 reported events.The device history records for the lot obtained through the ship history report (packaging lot) was reviewed.In addition, the corresponding lots for the purchased catheter tubing and connector were reviewed for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The recent angiodynamics complaint report was reviewed for the vortex ports product family and the failure mode "patient injury." no adverse trend was indicated.The reported complaint description cannot be confirmed given the nature of this failure mode.Potential contributing factors that may affect dvt/embolism include the following:high risk patient population, e.G.Oncology.Use of large fr catheter in smaller vessel.Port placement techniques, i.E.Excessive catheter manipulation that can damage the vessel wall.It was reported that the dual lumen port was used for apheresis red cell exchange therapy which is not an "indicated use" of the vortex port per the directions for use (dfu) the dfu includes the following guidance: "indications for use: the angiodynamics port line is indicated for any patient requiring repeated access of the vascular system for delivery of medications, nutritional supplementation, fluids, blood, blood products, and sampling of blood.Dual models are indicated for combination therapy, simultaneous infusions, withdrawal of body fluids, and bolus delivery during continuous infusion."contraindications: angiodynamics port systems should not be implanted in the presence of known or suspected infections, bacteremia, septicemia, and peritonitis, in patients who have exhibited prior intolerance to the materials of construction, or patients whose body size or tissue is insufficient to accommodate the size of the port or catheter.Potential complications: use of angiodynamics port systems involve potential risks normally associated with the insertion or use of any implanted device or indwelling catheter including but not limited to: infection; occlusion; thrombophlebitis, pneumothorax; catheter malposition; migration and inadequate anchoring; hemorrhage; vessel trauma, including puncture, laceration, and erosion of vessel and the skin; catheter pinch-off (compression of the catheter between the clavicle and the first rib); hematoma; clot formation; catheter fragmentation; embolization; cardiac arrhythmia; cardiac puncture; cardiac tamponade; fibrin sheath, endocarditis; implant rejection; thoracic duct injury; thromboembolism; peritonitis; thrombosis; and drug extravasation (leakage)." (b)(4).
 
Event or Problem Description
As reported by physician at hospital in (b)(6) utilizing vortex dual lumen port: "concerning the problems encountered with vortex in sickle cell patients: we have mainly a large problem of jugular thromboses and especially intra cardiac with a significant incidence of 7 thromboses for 18 vortex posed with 2 intracardiac thromboses that have been complicated by pulmonary embolism.These thromboses are most often asymptomatic and detected by trans-thoracic echocardiography and control echodoppler every 6 months in our unit.They occur in patients with an inflammatory profile and for the majority beyond one year of exposure." the ports were being used for apheresis which is not an approved application for this device.All thromboses were treated with xarelto.The port for this patient was removed and replace by another vortex port.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIODYNAMICS / VORTEX
Common Device Name
PORT & CATHETER, IMPLANTED
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key7556230
Report Number1317056-2018-00085
Device Sequence Number19441966
Product Code LJT
UDI-Device IdentifierH787LVTX52130
UDI-PublicH787LVTX52130
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K010767
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Catalogue NumberH787LVTX52130
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/07/2018
Initial Report FDA Received Date05/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-