MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIGMA HP UNI INS SZ4 7MM LM/RL; EARLY INTERVENTION : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIGMA HP UNI INS SZ4 7MM LM/RL; EARLY INTERVENTION : KNEE TIBIAL INSERT Back to Search Results
Catalog Number 102453407
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 05/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Der states that insert did not seem to engage properly.Second insert used in same manner seated properly.Three minutes surgical delay.Doi: (b)(6) 2018 left knee.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.Device history lot: product no.: 102453407 lot no.: 749079 device history review: no anomalies or deviations were found during the review of the warsaw site dhr.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP UNI INS SZ4 7MM LM/RL
Common Device Name
EARLY INTERVENTION : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7556586
Report Number1818910-2018-61006
Device Sequence Number2482082
Product Code HRY
UDI-Device Identifier10603295002581
UDI-Public10603295002581
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K070267
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 05/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Catalogue Number102453407
Device Lot Number749079
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/03/2018
Supplement Date Received by Manufacturer06/19/2018
Initial Report FDA Received Date05/31/2018
Supplement Report FDA Received Date06/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
-
-