Model Number M00562471
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/08/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Investigation results: visual evaluation of the returned device revealed that the snare loop was broken with evidence of being burned.Further examination noted that the handle cannula was found to be in good condition.A functional analysis was not performed due to the condition of the returned device.Due to anatomical/procedural factors encountered during the procedure; performance of the device was limited.Therefore, the most probable root cause classification is operational context.A review of the device history record (dhr) was performed and no deviations were found.A search of the complaint database revealed that no other complaints exist for the specified lot.
 
Event or Problem Description
It was reported to boston scientific corporation that a captiflex extra small oval snare was used during a colonoscopy procedure performed on (b)(6) 2017.According to the complainant, during the procedure, after the polyp was cut it was necessary to use cautery to stop the bleeding and it was noted that the wire inside the handle broke.The procedure was completed with another captiflex snare.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.This event has been deemed a reportable event based on the investigation results; loop broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIFLEX¿
Common Device Name
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
MDR Report Key7556597
Report Number3005099803-2018-01663
Device Sequence Number19453376
Product Code FDI
UDI-Device Identifier08714729019459
UDI-Public08714729019459
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K941750
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date05/10/2020
Device Model NumberM00562471
Device Catalogue Number6247
Device Lot Number20632795
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2018
Initial Report FDA Received Date05/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age80 YR
-
-