MAUDE Adverse Event Report: BOSTON SCIENTIFIC - CORK ROTALINK¿ PLUS; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - CORK ROTALINK¿ PLUS; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number H749236310060
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age at the time of event: 18 years or older.(b)(4).Device evaluated by manufacturer: the device was not returned for analysis.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
It was reported that the rotawire was moving together with the burr.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified right coronary artery.A 2.25mm rotalink¿ plus was selected for use.During procedure, after the first ablation, it was noted that the rotawire moved together with the burr.Both devices were removed as a system and a new rotawire was then advanced.It was observed that the same issue occurred.The rotalink plus was consequently removed and replaced with another of the same device.The procedure was then completed.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTALINK¿ PLUS
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI 
MDR Report Key7557595
Report Number2134265-2018-04921
Device Sequence Number10817434
Product Code MCX
UDI-Device Identifier08714729228394
UDI-Public08714729228394
Combination Product (Y/N)N
PMA/510(K) Number
P900056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/18/2019
Device Model NumberH749236310060
Device Catalogue Number23631-006
Device Lot Number0020899879
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2018
Initial Report FDA Received Date05/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-