| Model Number |
01-PA-65-45 |
| Medical Device Problem Code |
Disassembly (1168)
|
| Health Effect - Clinical Code |
Intraoperative Pain (2662)
|
| Date of Event |
05/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A device history review was conducted and the implant passed all specifications prior to being shipped.Complaint device was returned and inspected.The cause of the issue has not been able to be determined at this time.Further evaluation is being done and if cause is determined we will update this report with the results.
|
| |
|
Event or Problem Description
|
|
A posterior spinal fusion surgery was performed on the patient on (b)(6) 2018.Patient started to have pain a few weeks after this initial surgery.A ct scan was completed for this patient and the scan showed that the pedicle screw head disassembled.The patient was revised on (b)(6) 2018 and the disassembled screw was removed and replaced.
|
| |
|