Catalog Number 383408
Device Problem Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2018
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported before use, the nurse observed malformed tubing on the bd saf-t-intima¿ closed iv catheter system.No reports of serious injury or medical intervention noted.
 
Manufacturer Narrative
A sample was received for the purpose of our investigation.The investigators were able to observe the reported failure mode in the sample provided, the provided sample contained a transparent object cemented into the extension tube of the catheter adapter.The batch record also was reviewed, no related abnormalities were found.The most likely root cause of this failure mode is that the extension tubing came into contact with excess glue at the extension tube bonding station.The excess glue can occur, on occasion, during the cleaning of the glue gun.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SAF-T-INTIMA¿ CLOSED IV CATHETER SYSTEM
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer (Section G)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7557897
MDR Text Key109808577
Report Number3006948883-2018-00079
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K143610
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date07/06/2020
Device Catalogue Number383408
Device Lot Number7142060
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/22/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1