• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMMUCOR, INC. CAPTURE-R READY-ID; REAGENT RED BLOOD CELLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMMUCOR, INC. CAPTURE-R READY-ID; REAGENT RED BLOOD CELLS Back to Search Results
Catalog Number 0066214
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The implicated product had already expired before the customer had notified immucor of this event.Therefore, no direct product investigation could be performed.However, in lieu of a direct product investigation, immucor performed a dhr review, which yielded the following conclusions.The expected bulk manufacturing testing, using red blood cells from the bulk material, showed that the expected lea antigen positive red blood cell examples yielded the expected positive outcomes.The lea antigen is not routinely confirmed on the final microplate product as part of product final release.All other specifications were deemed to be acceptable, and the product was released to market.
 
Event or Problem Description
On (b)(6) 2018, a customer site reported an unexpected (b)(6) antibody identification when using capture-r ready-id when used on a galileo echo instrument, when tested on (b)(6) 2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURE-R READY-ID
Common Device Name
REAGENT RED BLOOD CELLS
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 5625
MDR Report Key7557993
Report Number1034569-2018-00134
Device Sequence Number1794516
Product Code KSZ
UDI-Device Identifier10888234001188
UDI-Public10888234001188
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
102707/0.0
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Medical Technologist
Type of Report Initial
Report Date (Section B) 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Medical Technologist
Device Expiration Date05/15/2018
Device Catalogue Number0066214
Device Lot NumberID357
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2018
Initial Report FDA Received Date05/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age29 YR
-
-