BOSTON SCIENTIFIC - MARLBOROUGH FLEXIVA (TM) ID HIGH POWER SINGLE-USE LASER FIBER; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M006R8403931
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2018 that a flexiva id 550 laser fiber was used during a bladder stone procedure performed on (b)(6) 2018.Reportedly, the laser unit used was lumenis p100 laser console with settings of 1.5 joules and 20 hertz.According to the complainant, about 15 minutes into the procedure, the operator attempted to advance the fiber in the working channel of the scope when the protective coating of the laser fiber sheared a little bit and the fiber body broke inside the working channel of the cystoscope.The laser fiber was taken out from the scope and scrape marks were noted inside the working channel of the cystoscope.The procedure was completed with another flexiva id 550 laser fiber.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIVA (TM) ID HIGH POWER SINGLE-USE LASER FIBER
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
MDR Report Key7558091
Report Number3005099803-2018-01686
Device Sequence Number1432802
Product Code GEX
UDI-Device Identifier08714729963806
UDI-Public08714729963806
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K100078
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/23/2020
Device Model NumberM006R8403931
Device Lot Number0000005234
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/10/2018
Initial Report FDA Received Date05/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-