• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 133MX TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 133MX TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number 133MX-15
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Device labeling: adverse reactions deflation- patients should be advised that the tissue expanders may deflate, and require replacement surgery.Deflation occurs when saline leaks through a damaged injection site or a damaged tissue expander envelope.
 
Event or Problem Description
Healthcare professional reported a right side deflation.Device remains implanted.
 
Event or Problem Description
Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified fold creases, wear abrasion and brown particles in device inner surface.A leak test was performed which identified no leakage.Based on the device analysis the final assessment is: no openings were found in the device.
 
Event or Problem Description
Healthcare professional reported a right side deflation.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 133MX TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key7558479
Report Number9617229-2018-03732
Device Sequence Number1818834
Product Code LCJ
UDI-Device Identifier10888628004207
UDI-Public10888628004207
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K862203
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/23/2021
Device Catalogue Number133MX-15
Device Lot Number3025145
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2018
Initial Date Received by Manufacturer 05/03/2018
Supplement Date Received by Manufacturer06/25/2018
08/18/2018
Initial Report FDA Received Date05/31/2018
Supplement Report FDA Received Date07/24/2018
09/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
-
-