Catalog Number CID-20-30
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that the stopcock on the sphere inflation device repeatedly popped off prior to patient contact.Another device was opened and the procedure was completed successfully.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the dimensional verification, complaint history, device history record, drawing, manufacturing instructions, quality control, specifications, and a visual inspection and functional evaluation of the returned device were conducted during the investigation.The complaint device was returned without a stopcock; accordingly, a representative device component was mated to the returned portion for the purposes of testing.The visual and functional inspections of the returned device noted no anomalies.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record for the complaint lot indicated that the device was manufactured under normal operating procedures.The device was sampled, inspected, and accepted per established specification.It should be noted there was one other reported complaint for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPHERE INFLATION DEVICE
Common Device Name
MAV SYRINGE, BALLOON INFLATION
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7558509
Report Number1820334-2018-01550
Device Sequence Number209691
Product Code MAV
UDI-Device Identifier10827002310275
UDI-Public(01)10827002310275(17)210228(10)96271388
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K953522
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCID-20-30
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 05/16/2018
Supplement Date Received by Manufacturer08/20/2018
Initial Report FDA Received Date05/31/2018
Supplement Report FDA Received Date08/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-