Model Number FAF07050
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number of the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during a shuntogram stent graft placement of the upper arm, the stent graft allegedly could not be fully deployed.It was further reported that the delivery system and stent graft were removed from the sheath without issue.The procedure was completed with another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
The device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode one flair endovascular stent graft was returned for evaluation.Visual inspection noted the stent was partially deployed.During guidewire patency test, the guidewire was advanced through the catheter with resistance past the incidental bends.The stent was also able to be fully deployed without issue.Therefore, the investigation is confirmed for partial deployment as the system was returned with the stent extending past the outer catheter.The root cause could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.The review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that during a shuntogram stent graft placement of the upper arm, the stent graft allegedly could not be fully deployed.It was further reported that the delivery system and stent graft were removed from the sheath without issue.The procedure was completed with another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLAIR ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7558997
Report Number2020394-2018-00769
Device Sequence Number13511078
Product Code PFV
UDI-Device Identifier04049519008790
UDI-Public(01)04049519008790
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P060002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/09/2018
Device Model NumberFAF07050
Device Catalogue NumberFAF07050
Device Lot NumberANCN3394
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2018
Initial Date Received by Manufacturer 05/04/2018
Supplement Date Received by Manufacturer09/17/2018
Initial Report FDA Received Date05/31/2018
Supplement Report FDA Received Date09/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-