| Model Number |
Z9002 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2018 and (b)(6) 2018.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that tecnis monofocal intraocular lens (model z9002 +18.0 diopter) was explanted from patient¿s left eye because of the patient experiencing residual refractive error.The lens was replaced with another z9002 iol but a lower diopter (+14.5).Patient reported as recovered.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: device available for evaluation ¿ yes, returned to manufacturer on 6/19/2018.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection at 10x microscope magnification was performed.Lens was returned cut in two pieces.Loose fibers/particles were observed on lens pieces related the handling of the unit out of a sterile environment.Residue of lubricant material was also observed on lens pieces.Both haptics was observed slightly distorted.The conditions of the lens returned is consistent with a unit that was previously handled and prepared for surgical process.The complaint issue reported could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no additional investigation requests for this order number have been received.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|