• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD ZILVER 635 BILIARY SELF-EXPANDING STENT; FGE CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ZILVER 635 BILIARY SELF-EXPANDING STENT; FGE CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number G36229
Medical Device Problem Codes Fracture (1260); Delivery System Failure (2905)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Us clearance # k040505 k163169.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.The investigation into this event is still being carried out.A follow up report will be submitted within the next 30 days with the investigation conclusions.
 
Event or Problem Description
Initial mdr is being submitted based on the device malfunction precedence: ¿flexor breaking during deployment (incl.Flexor/handle separation" and "stent fracture".As reported to customer relations: "while unsheathing catheter for stent deployment, outer catheter snapped, while inner portion of catheter remained intact.(catheter & unsheathing mechanism both flushed with saline prior to use).Stent kinked & then fell apart.Needed to use a new stent, which was deployed successfully." additional information provided by dm on (b)(6) 2018: "the midpart of the flexor sheath kinked, the physician could not deploy the stent inside the patient, the flexor sheath unraveled at the part of the kink (you will see this when the device is returned).".
 
Event or Problem Description
Mdr submitted based on the device malfunction precedence: ¿flexor breaking during deployment (incl.Flexor/handle separation" and "stent fracture" as reported to customer relations: "while unsheathing catheter for stent deployment, outer catheter snapped, while inner portion of catheter remained intact.(catheter & unsheathing mechanism both flushed with saline prior to use).Stent kinked & then fell apart.Needed to use a new stent, which was deployed successfully." additional information provided on 08may2018: "the midpart of the flexor sheath kinked, the physician could not deploy the stent inside the patient, the flexor sheath unraveled at the part of the kink (you will see this when the device is returned).".
 
Additional Manufacturer Narrative
Us clearance # k040505 k163169.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining (b)(4).Importer site establishment registration number: (b)(4).Problem statement as reported to customer relations: "while unsheathing catheter for stent deployment, outer catheter snapped, while inner portion of catheter remained intact (catheter & unsheathing mechanism both flushed with saline prior to use).Stent kinked & then fell apart.Needed to use a new stent, which was deployed successfully." additional information provided on 08may2018: "the midpart of the flexor sheath kinked, the physician could not deploy the stent inside the patient, the flexor sheath unraveled at the part of the kink (you will see this when the device is returned)." device evaluation the zib6-80-12.0-40 device involved in this complaint was returned for evaluation, without the original packaging.With the information provided, a physical examination and document based investigation was conducted.From customer testimony, it is known that the complaint device was advanced over a boston scientific amplatz wire guide.Resistance was not encountered during deployment.The patient¿s anatomy was not calcified or tortuous.Pre-dilation was conducted prior to stent deployment.The user confirmed that the handle was pulled towards the hub, and that the device was not pushed during deployment.The district manager confirmed that the target location was in the subclavian vein.The device related to this occurrence underwent a laboratory evaluation on the 31st may 2018.On evaluation of the returned device, it was observed that the outer flexor was separated, approximately 17cm from the connector cap.The stent was partially deployed, and 0.5cm was found to be exiting the flexor.The inner liner of the flexor was seen at the point of flexor separation.The red safety tab was not returned with the device.The stent was released in the lab, and was undamaged and 40mm in length.The device was flushed without issues, and a 0.035¿ diameter wire guide could pass through the device without resistance.The flexor length was 81cm, which was within specification of 80cm +1.3cm -0.8cm.There was no evidence of tactile damage on the other portions of the flexor.Complaint is confirmed as the failure was verified in the laboratory.It was found that the outer flexor was separated.Possible root causes for this occurrence could include the use of the device in a non-indicated location.From district manager's testimony, it is known that the complaint device was used in the subclavian vein.This could have created forces during deployment, that could have caused or contributed to the flexor separating.However, as the circumstances of use cannot be replicated in a laboratory environment, a definitive root cause cannot be determined.As per the product instructions for use: "the zilver 518 and 635 biliary stents are intended for palliation of malignant neoplasms in the biliary tree." document review prior to distribution all zilver biliary 635 devices are subject to visual inspection and functional checks to ensure device integrity as per fqc.A review of the relevant manufacturing records revealed no discrepancies that could have contributed to this complaint.Upon review of complaints, this failure mode has not occurred previously with this lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with this lot number.Summary complaint is confirmed as the failure was verified in the laboratory.It was found that the outer flexor was separated.The risk was determined to be no risk (category iii).According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.The procedure was completed with another stent.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZILVER 635 BILIARY SELF-EXPANDING STENT
Common Device Name
FGE CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
national technology park
limerick
MDR Report Key7559585
Report Number3001845648-2018-00250
Device Sequence Number2195184
Product Code FGE
UDI-Device Identifier10827002362298
UDI-Public(01)10827002362298(17)201117(10)C1420876
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG36229
Device Catalogue NumberZIB6-80-12.0-40
Device Lot NumberC1420876
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date05/04/2018
Event Location Hospital
Initial Date Received by Manufacturer 05/07/2018
Supplement Date Received by Manufacturer06/01/2018
Initial Report FDA Received Date06/01/2018
Supplement Report FDA Received Date06/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-