| Catalog Number |
7570855 |
| Medical Device Problem Code |
Device Slipped (1584)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor applicable imaging films were returned to manufacturer for evaluation therefore we are unable to determine the definitive cause of event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent 'percutaneous nail' surgery due to lumbar disc protrusion.Intra-op, the set screw slipped.The device came in contact with the patient.No patient complications were reported as a result of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: the center hex of the set screw is stripped.Per the reported though "according to the healthcare professional's opinion: the reported event is not related to use of medtronic products." if information is provided in the future, a supplemental report will be issued.
|
| |
|