MAUDE Adverse Event Report: MEDTRONIC IRELAND VALIANT CAPTIVIA - CW; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VALIANT CAPTIVIA - CW; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number VAMC3632C150TU
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Injury (2348)
Date of Event 05/11/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: other relevant device(s) are: product id: v amf3434c150tu, serial/lot #: (b)(4), (b)(6), (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A valiant stent graft system was implanted in a patient for the endovascular treatment of a ruptured 80mm diameter thoracic aortic aneurysm.It was reported that on ct four days later a type iii separation endoleak was observed between the proximal and distal stent graft.Another valiant stent graft was placed over the junction to resolve the endoleak one week later.As per the physician, the cause of the event was anatomy related due to the distal stent graft landing on extremely tortuous anatomy.No additional clinical sequelae were reported and the patient is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT CAPTIVIA - CW
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7560315
Report Number2953200-2018-00799
Device Sequence Number2044184
Product Code MIH
UDI-Device Identifier00643169346864
UDI-Public00643169346864
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/28/2019
Device Model NumberVAMC3632C150TU
Device Catalogue NumberVAMC3632C150TU
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2018
Initial Report FDA Received Date06/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-