| Catalog Number |
Y740D |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent gyn procedure on (b)(6) 2018 and suture was used.The suture broke when the surgeon pulled the suture through tissue a different type of suture was used to complete the procedure.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Pc-(b)(4).Date sent to the fda: (b)(4) 2018.Representative samples were returned for evaluation.The issued sample was not received for evaluation.During the visual inspection of the samples, no defects were found on the packages.Samples were opened and the swage and attachment area of needles were as expected.The sutures were dispensed without problems and examined along of the strand and no defects were observed.A functional test was performed and the tensile force were above the minimum requirements.Per the condition of the representative samples, no performance - breakage suture was found on the sutures and the tested sample met the finished goods requirements.
|
| |
|
Search Alerts/Recalls
|