MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 S-ROM*STEM BODY TRIAL 20; HIP INSTRUMENTS : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 S-ROM*STEM BODY TRIAL 20; HIP INSTRUMENTS : FEMORAL TRIALS Back to Search Results
Catalog Number 536320
Medical Device Problem Codes Bent (1059); Scratched Material (3020)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
New rep at (b)(6) medical center.Went through instrument inventory on (b)(6) 2018, and noticed a few instruments that were in an unacceptable condition.This instrument was not used during a surgery and no staff member had mentioned the damage.This is a srom 20 proximal body (color: blue).The groves where the neck trial sits are damaged ( bent, chipped).Asking that these are replaced and shipped to: (b)(6) medical center, united states.The damaged instruments will be returned to (b)(4), lot code: ah1105.P.S: on the complain tool; you are unable to add the lot code.The product won't add/upload when the lot is also typed.Patient consequence? :no.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary if information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
HIP INSTRUMENTS : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7560339
Report Number1818910-2018-61112
Device Sequence Number1529716
Product Code LXH
UDI-Device Identifier10603295172819
UDI-Public10603295172819
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number536320
Device Lot NumberAH1105
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2018
Initial Date Received by Manufacturer 05/02/2018
Supplement Date Received by Manufacturer06/22/2018
Initial Report FDA Received Date06/01/2018
Supplement Report FDA Received Date07/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-