PHILIPS MEDICAL SYSTEMS HEARTSTART MRX -EMS DEFIBRILLATOR Back to Search Results
Model Number M3536A
Device Problem Bent (1059)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Event Description
While at the customer site to evaluate another issue, the philips field service engineer (fse) observed bent pins on the battery pca.There was no patient involvement.
 
Manufacturer Narrative
One battery pca was returned for failure analysis.The reported problem was verified in the lab based on visual inspection.The problem was localized to j1 pin 6 which was found to be completely broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX -EMS DEFIBRILLATOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
nancy ataide
3000 minuteman road
andover, MA 01810
9786871501
MDR Report Key7560349
MDR Text Key109792483
Report Number1218950-2018-04671
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3536A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/14/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-