Pc-(b)(4).Date sent to the fda: (b)(4) 2018.A paper lid, a winding former with three re-parked needles from slot # 1 to 3 and 8, four needle/suture combinations form slot # 4 to 7 of product code (b)(4) were returned for analysis.This product code requires 8 strands per packet.During the visual inspection of sample, the swage and attachment area were as expected.Also, needle/suture combinations and suture pieces have begun with the process of disintegration, since the sutures were breakage in several sections.The product code (b)(4) contains an absorbable suture and as the sample was received open, the time of exposure of suture to the environment could not be determined and no functional test can be performed due to sample condition.In addition, the foil was not returned for evaluation to determine if any damaged was present, that could be attributable to degradation of the suture.Due to foil was not received for evaluation, no conclusion could be reached as to what may have caused the reported incident, the assignable cause of the reported complaint could not be determined.Representative samples were returned for evaluation.During the visual inspection of the samples, no defects were found on the packages.Samples were opened and the swage and attachment area of needles were as expected.The sutures were dispensed without problems and examined along of the strand and no defects were observed.A functional test was performed and the tensile force were above the minimum requirements.Per the condition of the representative samples, no performance - breakage suture was found on the sutures and the tested sample met the finished goods requirements.
|