MPRI SPRINT QUATTRO SECURE; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number 694758
Medical Device Problem Codes Signal Artifact/Noise (1036); Fracture (1260); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); Shock from Patient Lead(s) (3162)
Date of Event 04/30/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated oversensing associated with the right ventricular lead.Analysis of the device memory indicated the criteria for the right ventricular lead integrity alert were met.Analysis of the device memory indicated oversensing due to non-physiologic signals/sensing integrity counter.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient's right ventricular (rv) lead had triggered a lead integrity alert (lia) with increased sensing integrity counter (sic) and high rate non-sustained episodes.Additionally an alert for rv lead noise was also indicated.An intervention was completed and the low-voltage pace/sense portion of the rv lead was capped and new low-voltage pace/sense lead was implanted, while the high-voltage portion of the original lead remains in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was further reported that the right ventricular (rv) lead had developed a lead fracture which resulted in the patient receiving an inappropriate therapy.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7560419
Report Number2649622-2018-09163
Device Sequence Number2194132
Product Code LWS
UDI-Device Identifier00643169535961
UDI-Public00643169535961
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/03/2018
Device Model Number694758
Device Catalogue Number694758
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/07/2018
Supplement Date Received by Manufacturer05/07/2018
Initial Report FDA Received Date06/01/2018
Supplement Report FDA Received Date06/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DVBB1D1, ICD
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age83 YR
-
-