| Model Number |
694758 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Fracture (1260); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); Shock from Patient Lead(s) (3162)
|
| Date of Event |
04/30/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated oversensing associated with the right ventricular lead.Analysis of the device memory indicated the criteria for the right ventricular lead integrity alert were met.Analysis of the device memory indicated oversensing due to non-physiologic signals/sensing integrity counter.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right ventricular (rv) lead had triggered a lead integrity alert (lia) with increased sensing integrity counter (sic) and high rate non-sustained episodes.Additionally an alert for rv lead noise was also indicated.An intervention was completed and the low-voltage pace/sense portion of the rv lead was capped and new low-voltage pace/sense lead was implanted, while the high-voltage portion of the original lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was further reported that the right ventricular (rv) lead had developed a lead fracture which resulted in the patient receiving an inappropriate therapy.No further patient complications have been reported as a result of this event.
|
| |
|