| Model Number |
ENVEOR-N-US |
| Medical Device Problem Codes |
Difficult To Position (1467); Material Separation (1562); Detachment of Device or Device Component (2907); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: no product was returned. conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, during deployment, tension was felt in the delivery catheter system (dcs) and it was difficult to manipulate.An end cap separation was noted shortly after deployment was initiated.The valve and dcs were removed from the patient.When the dcs was inspected, a separation of the top of the capsule was noted and the valve was lodged inside of the dcs.Subsequently, another valve and dcs were used to successfully complete the procedure.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received that the loading was performed by experienced hospital staff with no misloads noted.The valve was recaptured one time for repositioning due to being too low.It was reported that the patient¿s common iliac arteries were very calcified.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: upon receipt at medtronic¿s quality laboratory, the valve was received loaded in the capsule of the delivery catheter system (dcs) with the capsule mostly closed.The device was returned with the end cap/screw gear snap fit connected.The handle and the tip-retrieval mechanism were intact.No damage was noted on the threading of the screw gear.The deployment knob was able to retract and advance the capsule.The trigger moved to fully advanced and retracted positions and locked in place when released.Voids were observed along the proximal end of the capsule.A break was observed in the capsule nitinol reinforcing frame near the proximal end of the capsule.The separation site was jagged and uneven.The valve could not be removed from the dcs.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.It was reported that the valve was recaptured once due to sub-optimal positioning.Recapturing was a feature of the evolut r system that allowed for additional attempts at accurately positioning the valve.Various factors can affect valve positioning including patient anatomy or physician technique, however the cause of the positioning difficulty could not be conclusively determined.The reported event indicated that the valve could not be deployed and it was noted that the end cap separated.End cap separation referred to an event where the end cap/screw gear snap fit fails and the capsule cannot retract.The failure occurred when the system forces exceed the tensile strength of the joint.When the valve and dcs were removed from the patient, a capsule separation was noted.Capsule separation typically occurred due to excessive compressive forces applied to the capsule.When a capsule separation occurred, the capsule may not retract to deploy the valve.Based on the information available, it cannot be determined if the reported unable to deploy event was caused by an end cap separation or by the capsule separation.Historically, a misloaded valve was a known cause of high forces leading to end cap separation.Misloads were also a known cause of the excessive compressive forces required to cause a capsule separation.No procedural images or fluoroscopic load check images were submitted for review, therefore it cannot be independently determined if the valve was correctly loaded onto the dcs.The force on the system is a cumulative effect that could have been increased by many factors, including patient anatomy.In this case, the iliac arteries were very calcified which may have contributed to the high forces in the system.A second valve was loaded onto another dcs and successfully implanted.No adverse patient effects were reported.There was no information to suggest a device quality deficiency that may have caused or contributed to this event, but a conclusive root cause cannot be determined at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|