• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. ARROW CVC KIT: 2-LUMEN 8 FR X 16 CM; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. ARROW CVC KIT: 2-LUMEN 8 FR X 16 CM; CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number AK-12802
Device Problems Bent (1059); Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/18/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The customer reports kinking and bending issue with the guide wire.
 
Manufacturer Narrative
(b)(4).The customer returned a spring-wire guide (swg) for evaluation.Visual examination of the swg revealed three bends in the body.No other defects or anomalies were found.The j-bend tip was undamaged and the distal and proximal welds were intact, full, and spherical.The three bends in the swg were located approximately 16, 23, and 219 mm from the proximal weld.The length and outer diameter of the swg were measured and were found to be within specification.A device history review was performed and did not reveal any manufacturing related issues.The instructions for use provided with the kit describes suggested techniques to minimize the likelihood of guide wire damage during use.It states that if resistance is encountered when attempting to remove the guide wire after catheter placement, the guide wire may be kinked about the tip of the catheter within the vessel.In this case it is recommended to withdraw the catheter relative to the guide wire about 2-3 cm and then attempt to remove the guide wire.The instructions caution that withdrawing the guide wire against the needle bevel or use of excessive force during removal could damage or break the wire.The reported complaint of guide wire kinking was confirmed by complaint investigation.The guide wire contained three bends.A device history record review was performed with no relevant findings.Based on the sample received, it was determined that operational context caused or contributed to this event.Teleflex will continue to monitor and trend for complaints of this nature.
 
Event Description
The customer reports kinking and bending issue with the guide wire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC KIT: 2-LUMEN 8 FR X 16 CM
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key7560516
MDR Text Key109773560
Report Number3003737899-2018-00060
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date06/30/2019
Device Catalogue NumberAK-12802
Device Lot Number13F18D0195
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2018
Date Manufacturer Received07/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-