• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-US XIA 3 TITANIUM ROD DIAM 6MM CP TI L 600MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US XIA 3 TITANIUM ROD DIAM 6MM CP TI L 600MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 48232600
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Injury (2348); No Code Available (3191)
Date of Event 05/10/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient was fixed from around t10 to the iliac bone.Post operatively, two rods broke on both sides near the iliac screw.Revision surgery was performed to remove the broken rods.Additional actions performed during revision: loosen the screw of s1 and replace to 1mm thick screw and fix it to the caudal side of the iliac connector with 4 rods.
 
Additional Manufacturer Narrative
The sales rep reported event of a was confirmed via correspondence with sales rep.The device was not returned, therefore visual, dimensional, functional analysis could not be performed.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.The surgical technique states "while the expected life of spinal implant components is difficult to estimate, it is finite.These components are made of foreign materials which are placed within the body for the potential fusion of the spine and reduction of pain.However, due to the many biological, mechanical and physicochemical factors which affect these devices but cannot be evaluated in vivo, the components cannot be expected to indefinitely withstand the activity level and loads of normal healthy bone.Delayed union or nonunion: internal fixation appliances are load sharing devices which are used to obtain alignment until normal healing occurs.In the event that healing is delayed, does not occur, or failure to immobilize the delayed/nonunion results, the implant will be subject to excessive and repeated stresses which can eventually cause loosening, bending or fatigue fracture.The degree or success of union, loads produced by weight bearing, and activity levels will, among other conditions, dictate the longevity of the implant.If a nonunion develops or if the implants loosen, bend or break, the device(s) should be revised or removed immediately before serious injury occurs." as it is unknown whether the patient fused, and the device was implanted for more than 2 years, the most likely root cause is fatigue fracture due to length of implantation.Bone quality and patient post op activity may have also contributed to the event.
 
Event or Problem Description
It was reported that a patient was fixed from around t10 to the iliac bone.Post operatively, two rods broke on both sides near the iliac screw.Revision surgery was performed to remove the broken rods.Additional actions performed during revision: loosen the screw of s1 and replace to 1mm thick screw and fix it to the caudal side of the iliac connector with 4 rods.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA 3 TITANIUM ROD DIAM 6MM CP TI L 600MM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key7560531
Report Number0009617544-2018-00123
Device Sequence Number11092339
Product Code NKB
UDI-Device Identifier04546540560582
UDI-Public(01)04546540560582
Combination Product (Y/N)N
PMA/510(K) Number
K113666
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number48232600
Device Catalogue Number48232600
Device Lot NumberMME
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/10/2018
Supplement Date Received by Manufacturer10/23/2018
Initial Report FDA Received Date06/01/2018
Supplement Report FDA Received Date11/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-