| Model Number |
ZCB00 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
04/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If implanted or explanted; give date: n/a (not applicable).The intraocular lens was removed (b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during implantation, the intraocular lens (model zcb00 +16.0 diopter) did not unfold evenly in the eye.The incision was enlarged and the lens was removed from the eye.No sutures were used.There was no patient injury.No additional information was provided to johnson and johnson surgical vision, inc.
|
| |
|
Additional Manufacturer Narrative
|
|
Per follow-up information, there was a problem with the trailing haptic getting stuck in the cartridge causing the lens to become everted in the bag based on the customer's chart.The physician was unable to reposition it properly and it was unstable.He decided to remove the zcb00 lens and implant a 3-piece lens instead.Incision was slightly widened to accommodate removal of zcb00 lens and insertion of the 3-piece lens b&l li61a0 14.0.There was no vitrectomy performed on os (left eye).Device evaluation: the product testing could not be performed as the product was not returned.The customer's reported complaint cannot be confirmed.Manufacturing record review: a manufacturing record review could not be performed as no serial number was provided.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: initial supplemental report inadvertently reported cartridge investigation results.The complaint was for the lens; and the lens investigation results are provided in this supplement.Device available for evaluation? yes.Returned to manufacturer on: 06/06/2018.Device returned to manufacturer? yes.Device evaluation: the lens was received.Only half of the lens was returned.Based on the condition of the lens, the reported complaint could not be confirmed and product deficiency could not be identified.Manufacturing record review: manufacturing record review of the production order and related documents revealed that the product was manufactured and released according to specification.There was no discrepancy found during the review.A search revealed that no other complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|