| Model Number |
2205222 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer, at this time, for evaluation.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
|
| |
|
Event or Problem Description
|
|
It was reported that while infusing a high concentration vitamin through the device, leakage was confirmed from the connection part of the device and the administration set.No patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial complaint appeared to be a reportable occurrence.Once the sample was returned and evaluated it was determined that a reportable event had not occurred.The returned sample, with a non-bard administration set, exhibited no evidence of the alleged event.The event is unconfirmed per the investigation results.Since no device malfunction occurred and there is no allegation of a serious injury to a patent, user, or other person, this event is deemed not reportable per 21 cfr part 803.
|
| |
|
Event or Problem Description
|
|
It was reported that while infusing a high concentration vitamin through the device, leakage was confirmed from the connection part of the device and the administration set.No patient harm reported.
|
| |
|