• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS WINGED INFUSION SET WITH INJECTION SITE 22G X 0.75IN; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS WINGED INFUSION SET WITH INJECTION SITE 22G X 0.75IN; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number 2205222
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer, at this time, for evaluation.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
It was reported that while infusing a high concentration vitamin through the device, leakage was confirmed from the connection part of the device and the administration set.No patient harm reported.
 
Additional Manufacturer Narrative
The initial complaint appeared to be a reportable occurrence.Once the sample was returned and evaluated it was determined that a reportable event had not occurred.The returned sample, with a non-bard administration set, exhibited no evidence of the alleged event.The event is unconfirmed per the investigation results.Since no device malfunction occurred and there is no allegation of a serious injury to a patent, user, or other person, this event is deemed not reportable per 21 cfr part 803.
 
Event or Problem Description
It was reported that while infusing a high concentration vitamin through the device, leakage was confirmed from the connection part of the device and the administration set.No patient harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WINGED INFUSION SET WITH INJECTION SITE 22G X 0.75IN
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7561109
Report Number3006260740-2018-01176
Device Sequence Number2063180
Product Code FMI
UDI-Device Identifier00801741065873
UDI-Public(01)00801741065873
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K863606
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2205222
Device Catalogue Number2205222
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2018
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer06/18/2018
Initial Report FDA Received Date06/01/2018
Supplement Report FDA Received Date07/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-