BOSTON SCIENTIFIC - GALWAY EXPRESS® LD ILIAC / BILIARY; STENT, ILIAC Back to Search Results
Model Number H74938046830750
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr: a visual and microscopic examination was performed on the returned device.The sheath used during the procedure was not returned for analysis.A stringy clear material was noted to be attached to the stent on the express ld device.The express ld device was returned with the stent fully mounted onto the delivery device.The balloon was tightly folded and the stent was in the correct position on the balloon material.No damage or any issues were noted on the returned delivery device that could have contributed to the complaint incident.The stent was returned fully mounted onto the delivery system.A visual and microscopic examination identified a clear stringy foreign material intertwined around some stent struts approximately 4mm distal to the distal edge of the proximal markerband.The clear stringy foreign material was noted to be approximately 240mm in length.No issues were noted with the returned stent that could have contributed to the complaint incident.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is caused by other device as another device/drug/subsequent procedure caused the complaint event.(b)(4).
 
Event or Problem Description
It was reported that a foreign matter was found stuck on the stent.The target lesion was located in the superior mesenteric artery.An 8.0x30x75cm express® ld iliac / biliary stent was selected for use to treat the lesion.However, it was noted that there was a string hanging off of the stent.The procedure was completed using a different device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS® LD ILIAC / BILIARY
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7561184
Report Number2134265-2018-04767
Device Sequence Number16500021
Product Code NIO
UDI-Device Identifier08714729579670
UDI-Public08714729579670
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P090003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/10/2019
Device Model NumberH74938046830750
Device Catalogue Number38046-83075
Device Lot Number0019797476
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2018
Initial Report FDA Received Date06/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
-
-