| Model Number |
8253200RF |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Continuation of medical devices: information references the main component of the system.Other relevant device(s) are: product id: 8253001rf, serial/lot #: (b)(4), ubd: n/a, udi#: (b)(4), 510k number - k083124.(b)(4) (nim 3.0 patient interface): no product analysis results available.Device evaluation anticipated but not yet begun.(b)(4) (nim 3.0 mainframe): the product analysis indicates that the reported issue could not be duplicated.A 24-hour burn-in testing was performed.There was no fault found with the device.The device tested and passed all manufacturing specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The customer reported that the device was working intermittently.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
The product analysis indicates the patient interface came in with broken cable clip and patient interface pcb failure.The reported issue was confirmed.The p/i pcb board was replaced due to failure.The broken cable clip was also replaced.The patient interface was repaired and successfully tested to specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|