HOLOGIC, INC. NOVASURE IMPEDANCE CONTROLLED EA SYSTEM; UTERINE ABLATION DEVICE Back to Search Results
Model Number NS2000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Uterine Perforation (2121)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not being returned therefore, a failure analysis of the complaint device cannot be completed.Device history record (dhr) review was conducted for the reported identification number.The lot was released meeting all qa specifications.If additional relevant information is received, a supplemental medwatch will be filed.According to the instructions for use (ifu) warnings: use caution not to perforate the uterine wall when sounding, dilating, or inserting the disposable device.If the disposable device is difficult to insert into the cervical canal, use clinical judgment to determine whether or not further dilation is required.The novasure system performs a cavity integrity assessment (cia) test to evaluate the integrity of the uterine cavity, and sounds an alarm warning of a possible perforation prior to treatment.(b)(4).
 
Event or Problem Description
It was reported the physician performed a novasure endometrial ablation on (b)(6) 2018 and the physician noted a perforation.It is unknown if intervention was required.We have been unable to obtain additional information surrounding this event.
 
Event or Problem Description
Note: this report pertains to the second of two hologic devices used in the same procedure.See associated medwatch, manufacturer's report# 1222780-2018-00151.On 06/18/2018, it was reported there was no intervention to the patient and the patient is doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVASURE IMPEDANCE CONTROLLED EA SYSTEM
Common Device Name
UTERINE ABLATION DEVICE
Manufacturer (Section D)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
MDR Report Key7562086
Report Number1222780-2018-00113
Device Sequence Number19381805
Product Code MNB
UDI-Device Identifier45420045501134
UDI-Public(01)45420045501134(10)17K18R
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P010013
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date11/18/2018
Device Model NumberNS2000
Device Catalogue Number915002-01
Device Lot Number17K18R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2018
Supplement Date Received by Manufacturer06/18/2018
08/05/2018
Initial Report FDA Received Date06/01/2018
Supplement Report FDA Received Date06/27/2018
08/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RADIO FREQUENCY CONTROLLER: SERIAL NUMBER UNKNOWN; RADIO FREQUENCY CONTROLLER: SERIAL NUMBER UNKNOWN
Outcome Attributed to Adverse Event Other;
-
-