Sc-1132 (sn: (b)(4)) device evaluation indicated that the reported complaint of ipg stopped working was confirmed.Upon receipt, the device was not functional, subsequently, the device was cut open and the internal inspection revealed excessive sleep current leakage.A hot spot was observed on the component (33.8°c).The impedance between vh and ground was low.These are the typical symptoms of the asic damage due to exposure to high-voltage transients, causing internal shorts in the component.It was reported that the device may have damaged during a procedure to deepen the ipg pocket.
|