| Catalog Number |
251710706 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
05/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that attune knee revision rp insert trial sz 7 6mm did not fit in trial femur.Manufacturing error.A replacement trial was ordered and put our set, and works well.Might be a problem with a specific lot.2517-10-706 lot az887033.Surgical delay for 5 minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # pc-(b)(4).Investigation summary evaluation and measurement of the returned device was unable to confirm the reported event.The returned device met print specification.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|