CLEARSTREAM TECHNOLOGIES LTD. SLEEK RX PTA CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number 425-2010X
Medical Device Problem Codes Inflation Problem (1310); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during a right lower extremity arteriogram, the balloon allegedly failed to inflate.After multiple inflation attempts, the hub was allegedly loose from the catheter and disconnected from the catheter during removal.The procedure was completed with another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLEEK RX PTA CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford
EI 
MDR Report Key7562526
Report Number9616666-2018-00070
Device Sequence Number13835307
Product Code LIT
UDI-Device Identifier00801741132254
UDI-Public(01)00801741132254
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K072947
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2020
Device Model Number425-2010X
Device Catalogue Number425-2010X
Device Lot NumberCMBR0125
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/07/2018
Initial Report FDA Received Date06/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-