| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Codes |
Tingling (2171); Dizziness (2194)
|
| Date of Event |
05/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low blood glucose results accompanied by symptoms.The customer is concerned with results obtained of 92mg/dl (non fasting).The expected fasting blood glucose test result range is 129 to 140mg/dl or 170 to 180mg/dl non-fasting.The customer did report symptoms of tingling feet and feeling lightheaded.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the living room.During the call on (b)(6) 2018, a back to back blood test was performed non-fasting by the customer and produced test results of 99 and 97mg/dl using true metrix meter.The test strip lot manufacturer's expiration date is 08/21/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6).
|
| |
|