| Model Number |
R SERIES |
| Medical Device Problem Codes |
Loss of Power (1475); Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed self-test and inappropriately shut down.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The complainant was contacted for return of the device.The device has not been returned to zoll for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported malfunction was observed during review of the activity logs.However, the reported problem could not be duplicated with the device.The pace/defib engine board was replaced as a precaution the device was recertified and returned to the customer.The electrode pads were not returned to zoll medical corporation for evaluation.No trend is associated with reports of this type.
|
| |
|