Catalog Number 442194
Health Effect - Clinical Code Death (1802)
Date of Event 04/10/2018
Type of Reportable Event Death
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Investigation: bd diagnostic systems (bdds) quality investigated the complaint of a potential false positive biofire test result when used with bd bactec¿ peds plus¿/ f culture vial.The biofire bcid identified proteus in a positive patient blood culture bottle with the results being reported out after the patient expired.The patient was 3 months in age and had been in the nicu for an infection since birth.The caregivers were aware that the patient had an e.Coli infection prior to receiving the biofire results.The patient was treated with gentamicin and vancomycin.Proteus did not grow in culture nor was it seen on gram stain.According to the hospital the patient¿s death was not due to any changes in treatment.Bdds investigation findings to date are as follows: we have tested more than 400 samples for bioburden and/or molecular testing.No viable proteus organisms have been recovered in any sample including raw materials, in-process water, the deionized (di) water system, and the environmental samples throughout the manufacturing process.No proteus dna has been detected by molecular testing with biofire filmarray bcid assay in the water supply, the di water system and the environmental samples throughout the manufacturing process.Consistent with customer reports, some lots of final product tested positive for proteus sp with biofire filmarray bcid assay.There have been no recent changes to raw materials, design, manufacturing process, or distribution practices.The remainder of bdds investigation will focus on the following areas: additional testing of the dehydrated media and the raw material components.Cross-reactivity evaluation of purified cultures of environmental organisms found during bioburden testing.Testing of additional in-process samples.Further actions are to be determined based on the remaining investigation results.
 
Event or Problem Description
It was reported that biofire identified proteus in a bd bactec¿ peds plus¿/ f culture vial with results being reported out after the patient expired.The patient was three months in age and had been in the nicu for the infection since birth.The caregivers were aware that the patient had e.Coli prior to receiving the biofire results.The patient was treated with gentamicin and vancomycin.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BACTEC¿ PEDS PLUS¿/ F CULTURE VIALS
Common Device Name
CULTURE VIAL
Manufacturer (Section D)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
MDR Report Key7562714
Report Number2647876-2018-00017
Device Sequence Number13503969
Product Code MDB
UDI-Device Identifier00382904421946
UDI-Public00382904421946
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K954927
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Expiration Date10/31/2018
Device Catalogue Number442194
Device Lot Number8015973
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/10/2018
Initial Report FDA Received Date06/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age3 MO
-
-