| Model Number |
ASU1201 |
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
05/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Information has been forwarded to the manufacturing facility for investigation.Results pending at this time a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
Continuous suction without pressing the suction piston on the trumpet valve, the red control button didn't work.
|
| |
|
Additional Manufacturer Narrative
|
|
No sample was received for evaluation therefore we are unable to confirm the issue reported.However, cardinal health has become aware of an issue in which the hydroline and pulse wave trumpet valves may not close properly, which could cause continuous suction.There is a remote possibility that inability to control suction during a procedure may result in loss of carbon dioxide in the abdomen, reduced visibility of the surgical site and potential injury to organs and body tissue.Cardinal health has issued a voluntary recall that was initiated on june 11th 2018.
|
| |
|